Calibration

GB
Select the Calibration menu as described
on page 11.
1. Press Enter.
2. Select:
Calibrate
Calibration Coefficient
Calibration Time or
Password Protected Calibration
3. Press Enter. If requested — enter pass-
word, see page 29.
4. Set the desired value using + and –.
Use > to move to next character and after
the last character.
(5. Press Undo to delete the setting and
revert to the previous value.)
6. Press Save to save the set value.
Revert to Detection Mode/Analysis
Mode/Combined Mode by pressing
Esc twice.
Explanations
Calibration Time
The number of seconds that measurement is in progress when calibrating in
the Analysis Mode. The default value is 10 seconds but values from 5 to 30
seconds can be used.
Calibration Coefficient
Calibration parameter. See page 30.
Password Protected Calibration
The calibration function can be password protected so that calibrating cannot be
performed by unauthorised users. Note! Factory default is no password.
1
Adixen Scandinavia - User's manual H2000 PLUS
Last successful calibration
Calibration
Calibrated
00-0-
11:0:
Esc
<
2
Calibration
Calibration Coefficient
1.00E+01
<
+
Calibration
Calibration Coefficient
.00E+01
Undo
5
>
Enter
3
>
4
Save