A typical set-up procedure for the Detec-
GB
tion Mode is:
• Set up a reference leak which corresponds
to the smallest leak you wish to detect.
• Put the probe close to the reference leak
and note approximately what reaction you
get (no reaction, small, medium, high, full
scale) within the first few seconds.
• Set the sensitivity. This can be done
permanently under the menu Detection
Mode Settings or temporarily as a
Direct Sensitivity Adjustment on the

Adixen Scandinavia - User's manual H2000 PLUS
display (unless you have set this function
to OFF under menu Detection Mode
Settings. See pages 17 and 18). There is
also an Autoranging function which can
be selected under the Sensitivity Settings
menu.
Note! If the Detection Mode is used and
the alarm function is required to be activa-
ted at a particular calibrated level, then the
unit must be calibrated in accordance with
the instructions on page 14 and 29. The
reason for this is that the alarm is always
based on the Analysis Mode even when
the Detection Mode is displayed.