Use As Smoke Generator - MSA TOXIMETER II Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 23
GB
GB
Fig. 4: Adjusting the
number of pump strokes
Fig. 5: Adjusting the factor
Fig. 6: Insertion into outlet
side
Fig. 7: Insertion into inlet
side
The port for insertion of the tubes must be
examined. Failure to do so will result in
destruction of the membrane of the pump
and a faulty measurement.
The pump port are clearly marked with a
blue or white directional arrow. The
label on the detector tube package
identifies the tube port to be used by a
coloured arrow.
During correct use of the detector tube the
arrow on the port is in same direction as
the arrow on the tube.
Break off both tips and insert firmly the
tube into the inlet or outlet port.

6 Use as Smoke Generator

Smoke generator tubes and cartridges to be
inserted only on the outlet port of the pump.
Use of smoke generating tubes or
cartridges on the inlet port of the pump
will destroy the membrane pump!
Break off both tips and insert firmly the tube into
the outlet port (blue arrow). To terminate smoke
generation remove the smoke tube from the port
and/or switch off pump.