Safety
For maximum safety, observe the following warnings and cautions before using accessories and peripherals
with the SonoSite SII ultrasound system and transducers.
Electrical safety
To avoid the risk of electrical shock and possible fire:
Do not plug the power cord of the SonoSite SII stand into a multiple portable
Use only properly grounded equipment. Shock hazards exist if the power
WARNINGS:
Do not connect more than one peripheral at a time.
Inspect the power supply, plugs, cables, and power cords on a regular basis for
Use only accessories and peripherals recommended by FUJIFILM SonoSite,
Equipment safety
To avoid tipping, and the risk of personal injury or damage to the system, do not
WARNING:
move the stand using the handle on the front of the ultrasound system.
SonoSite SII Stand User Guide
socket outlet (MPSO); do not use an extension cord.
supply is not properly grounded. Grounding reliability can only be achieved
when equipment is connected to a receptacle marked "Hospital Only, "
"Hospital Grade, " or the equivalent. The grounding wire must not be removed
or bypassed.
damage.
including the power supply. Connection of accessories and peripherals not
recommended by FUJIFILM SonoSite could result in electrical shock. Contact
FUJIFILM SonoSite or your local representative for a list of accessories and
peripherals available from or recommended by FUJIFILM SonoSite.
13