Warning And Safety Information - Sirona NITRASEAL Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1 Warning and safety information

1
Warning and safety information
Intended use
General recommendations
4
Operating Instructions NITRASEAL
Even when it is not strictly necessary, the instrument
wrapping indicates a special care paid to the hygiene
and sterilisation and it is especially effective for a proper
infection safety.
Wrapping the instruments means to preserve their ste-
rile condition also during the period following the sterili-
sation and until their use on the patient.
A properly sealed envelope makes up a protection bar-
rier that prevents contamination from the external envi-
ronment.
For this to be true, it is necessary to use envelopes that
have been properly sealed.
The NITRASEAL Thermal Sealer guarantees the best
efficiency and reliability for the sealing system, avoiding
the risk of overheating and burning.
Inexperienced staff can easily use this equipment with a
quick training.
In order to maintain the instruments for a long time and
avoid problems it is recommended to read the following
instruction carefully.
1.
Wrap the instruments one by one.
2.
If more instruments must be wrapped together make
sure to place instruments of same material (on the
contrary, dark spots could appear at the contact
point due to electrolysis).
3.
It is recommended to use good quality paper and
propylene material. The paper surface (more po-
rous) must allow the vapour exit. The NITRASEAL
envelopes has been widely tested in order to give
the best results both for quality and durability.
4.
For the sterilisation of wrapped instruments, it is ne-
cessary to use longer sterilisation times. The vapour
must actually filter within the envelopes and this re-
quires longer times.
5.
Most autoclaves normally require a time of 12 minu-
tes at a temperature of 134 °C and 30 minutes at
121 °C.
6.
The envelopes should be placed inside the autocla-
ve with the paper surface up. This allows the vapour
exit easier during the drying phase.
7.
Do not overload the sterilisation chamber! Leave a
space for steam circulation.
8.
In case of sterilisation of wrapped hand pieces and
turbines, try to arrange the envelopes so that the
head of the hand piece is turned towards the upper
Sirona Dental Systems GmbH
61 79 761 D 3367
D 3367.201.02.01.02

Publicidad

Tabla de contenido
loading

Tabla de contenido