Section 1
•
Select the radio button for desired maximum rate
•
Type password "ACCELERATE10K" in the password field.
•
To activate click [Apply to Stimulator]
A message will be displayed stating that all programs will be cleared from the
stimulator.
•
Click [Yes].
A message will be displayed stating that the Stimulator maximum rate has been
updated,
•
To exit, click [Close].
When programming a patient at 1,200 Hz or below first set the Maximum Rate to 1,200 Hz
using Bionic Navigator HR 1.0 Software, and then use Bionic Navigator 1.2 Software to
program the patient.
When programming a patient above 1,200 Hz first set the appropriate Maximum Rate
using Bionic Navigator HR 1.0 Software and then program the patient with the same
software.
System
Click [System Info] to check stimulator and Remote Control information.
•
To manually create a separate compressed file of database and logs, click
[Backup Data/Logs].
To check stimulator and Remote Control information regarding
identification, type, version, battery voltage and status, click [System Info].
To display relative lead positions, click [EGL Scan].
Backup Data/Logs
To create a compressed file of database and logs.
•
Insert [Removable media] to copy database and logs.
- This Removable media should not be labeled "CLINIC"
•
Click on Backup Data/Logs button
- Database is successfully copied when message appears
•
Click [OK] to exit.
System Information
•
System Information is only available when a stimulator is connected to the
Clinician's Programmer.
EGL Scan™
The EGL Scan image is a display of relative lead positions.
•
Click [EGL Scan] to access image.
- EGL Scan is available only if supported by the stimulator.
- Use of EGL Scan with an external trial simulator will raise a prompt for
a patch electrode. The prompt should be ignored.
Optionally, select the number of samples and the amplitude to be used for EGL
•
Scan before initiating EGL Scan.
•
To print EGL Scan image, click [Print] or [Cancel] to exit.
Additional Information
To change patient settings use [Patient Options].
Bionic Navigator HR 1.0 Software Guide
Clinician Programmer
90868740-07 Rev A 3 of 84