Section 4
Section 4
Patient Hardware

Links and Resets

Description
Linking is the Boston Scientific term for programming the Remote Control to
communicate with a specific stimulator, and determining the therapy program the
devices will share. Resetting describes the process of clearing previously stored
programs from a clinic Remote Control and /or Trial Stimulator.
Operation
A new Remote Control may be linked to the new Trial Stimulator, loaded with
a preliminary "program" through the Clinician Programmer during the surgery
and delivered to the patient.
A Remote Control linked to a Trial Stimulator but kept at the clinic must be
de-linked from the Trial Stimulator before it can be used with another Trial
Stimulator or with an IPG. While linking with the new device, the remote will
prompt your decision concerning how to reset the equipment.
Remote Control
Linking an unlinked Remote Control will automatically display its status when any
P
button is pressed. Press Program
on the remote to initiate a link with a stimulator
within telemetry range:
The remote will identify the targeted device and determine the stimulator's
programmed status.
The remote will compare its programmed status against that of the stimulator and
prompt you to choose either program set, if applicable.
If the preferred program set is the Remote Control programs, you will need to
enter the clinician password before downloading programs from the remote.
If neither the remote nor the stimulator has stored programs, device linking is
immediate.
De-linking
P
Press and hold the Area
and Program
"
approximately three seconds to access the Enter Clinician Options screen.
From the Enter Clinician Options screen, select "To Clear Link".
At the prompt, enter the clinician password.
The unlinked Remote Control will immediately display its status. You may
re-link immediately, or press
"
Resets from the Clinician's Programmer
Plug the IR Dongle/USB Extension into the USB port and slip Remote Control
into the slip pouch.
Position the IR window of the Remote Control in-line with the IR window of the
dongle.
Enter the [Clinician's screen] if appropriate on the Remote Control.
A message "CP Ready" will be displayed.
buttons on the Remote Control for
to postpone linking.
Patient Hardware
Bionic Navigator HR 1.0 Software Guide
90868740-07 Rev A 7 of 84