Section 7
Section 7
Testing

Impedance

Description
The Impedance button is used to verify electrical integrity. Lead impedance is measured
and displayed for each of the IPG's 16 electrode contacts. Impedances over 4500 Ω
are considered to be resultant from open or unconnected wires, displayed with an X.
Access/Exit
To access lead impedance, click [Ω] from the Programming Screen or EGL
Scan Screen.
To perform the measurement again, click [Measure].
To exit, click [Close].
Additional Information
Even though an X indicates higher than normal impedances, electrodes may still
be included in stimulation configuration.
All measured impedances are saved into the database.
Testing
Bionic Navigator HR 1.0 Software Guide
90868740-07 Rev A 11 of 84