3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Calibration:
Automatically with up to three buffer solutions, sequence during calibration optional, freely
definable buffers can be input.
Default buffer solutions according to DIN 19 266 and NBS, or technical buffers:
pH = 1.00; pH = 4.00; pH = 4.01; pH = 6.87; pH= 7.00; pH = 9.18; pH = 10.00
Input:
Measurement input 1: pH/mV-input with electrode socket according DIN 19 262/or BNC
Measurement input μA: (Dead-Stop-) connector for double platinum electrode
Measurement input Pt 1000: Temperature sensor probe for resistance thermometer Pt 1000
Power supply:
power supply 100-240 V; 50/60 Hz, power input: 30 VA
RS-232-C Interface:
separated galvanically through photocoupler, Daisy Chain function available
Data bits:
Stop bit:
Start bit:
Parity:
Baud rate:
Address:
RS-232-1
for computer, input Daisy Chain
RS-232-2
devices of SI Analytics
- Titrator TitroLine
- Sample Changer TW alpha plus, TW 7400
- Piston burette TITRONIC
- Balances of the types Mettler, Sartorius, Kern, Ohaus, (for more, please contact us)
- Exit Daisy-Chain
USB Interface:
2 x USB-type-A and 1 x USB-type-B
(„master") for connecting of USB keyboard, - printer, - manual controller, - data media
USB-type A
(e.g. USB stick) and USB-Hub
(„slave") for connecting a PC
USB-type B
Ethernet Interface:
for connecting a local network (LAN)
Stirrer/pump:
12V DC out, 500 mA
power supply for stirrer TM 235 and KF titration stand TM 235 KF
Housing:
Material:
Polypropylene
Front keyboard: polyester coated
Dimensions:
15.3 x 45 x 29.6 cm (W x H x D), height incl. interchangeable unit
Weight:
approx. 2.3 kg for basic unit
approx. 3.5 kg for complete device incl. interchangeable unit (with empty reagent bottle)
Ambient conditions:
Climate:
Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °C
(Connection sockets: 2 x 4 mm)
(Connection sockets: 2 x 4 mm)
Use the power supply TZ 1853 only!
adjustable, 7 or 8 Bit (default: 8 Bit)
adjustable, 1 or 2 Bit (default: 1 Bit)
static 1 Bit
adjustable: even / odd / none
adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
adjustable, (0 to 15, default: 01)
®
:
®
7000 / 7500 / 7500 KF / 7750 / 7800
®
Do not use the device in hazardous locations!
300 and 500, TITRONIC
®
®
110 plus, TITRONIC