130
Interchangeable units:
Compatibility:
units are compatible to the:
- the titrators TitroLine
- the Piston Burette TITRONIC
Recognition:
automatically through RFID
Recognition of unit size and characteristics of the Titration- or dosing solution
Valve:
volume neutral cone valve made from fluorocarbon polymers (PTFE), TZ 3000
Cylinder:
borosilicate glass 3.3 (DURAN
Hoses:
FEP hose set, blue
Bracket for
supply bottle:
suitable for square glass bottle and misc. reagent bottles
Materials:
borosilicate glass DURAN
Dimensions:
15,3 x 45 x 29,6 cm (W x H x D) incl. reagent bottle
Weight:
approx. 1.2 kg for interchangeable unit WA incl. empty reagent bottle
Dosing accuracy:
after DIN EN ISO 8655, part 3:
Accuracy:
Precision:
(in dependence of the used interchangeable unit)
Dosing accuracy of the Titrator TitroLine
Interchangeable
Volume
unit type No.
[ml]
WA 05
WA 10
10.00
WA 20
20.00
WA 50
50.00
®
7000 / 7500 / 7500 KF / 7750 / 7800
0.15 %
0.05 - 0.07 %
®
Tolerances of the Ø
of the glass cylinder [mm]
5.00
± 0.005
± 0.005
± 0.005
± 0.005
®
500
®
)
®
, fluorocarbon polymers (PTFE), stainless steel, polypropylene
7000 with interchangeable units (WA):
Dosage error according to
i
100 % volume [%]
Reproducibility [%]
± 0.15
± 0.15
± 0.15
± 0.15
0.07
0.05
0.05
0.05