Resperation Rate Accuracy
Accuracy data was obtained from a comparison of a manual count of respiratory rate in healthy adult subjects and in hospitalized pediatric
subjects for whom post-operative monitoring of respiratory rate was indicated.
Respiration rate accuracy for the rainbow Acoustic Monitoring sensor and instrument has been validated for the range of 4 to 70 breaths per
minute in bench top testing. Clinical validation was performed with the RAS-125 and RAS-125c sensors and instrument for up to 30 breaths per
minute in adult subjects >30 kg, and up to 50 breaths per minute in hospitalized pediatric subjects, age 2 years or older and >10 kg.
+70 C
Body Weight
-40 C
+1060 hPa - +500 hPa
795 mmHg - 375 mmHg
Application Site
NON
STERILE
5%-95% RH
Breaths per Minute, Accuracy Range
LATEX
95%
The following plots graph the results for the healthy adult subjects vs. the control method:
%
5%
RAS-125
Adult > 30 kg
Neck
4 to 70 ± 1 breath per minute
40
100 ≤ Number of Samples
10 ≤ Number of Samples < 100
0 < Number of Samples < 10
35
30
25
20
15
10
5
5
10
Healthy Adult Subjects: RAS-125 vs. Control (manual count)
40
1000 ≤ Number of Samples
100 ≤ Number of Samples < 1000
10 ≤ Number of Samples < 100
30
0 < Number of Samples < 10
20
10
0
−10
−20
−30
5
10
)RAM + Control)/2 )Respiration per Minute)
Healthy Adult Subjects: RAS-125 vs. Control (manual count)
RAS-125c
Adult, Pediatric > 10 kg
Neck
4 to 70 ± 1 breath per minute
15
20
25
30
Control )Respiration per Minute)
Bias=0.18, STDEV=1.31
15
20
72
35
40
25
8995B-eIFU-1115