METTLER TOLEDO ASSUMES NO RESPONSIBILITY FOR CORRECT INSTALLATION
OF THIS EQUIPMENT WITH A DIVISION 2 OR ZONE 2/22 AREA. THE INSTALLER
MUST BE FAMILIAR WITH ALL DIVISION 2 OR ZONE 2/22 WIRING AND
INSTALLATION REQUIREMENTS.
EARLIER MODELS OF THE PANTHER TERMINAL THAT ARE NOT MARKED
(FACTORY-LABELED) AS DIVISION 2 OR EUROPEAN CATEGORY 3 APPROVED
MUST NOT BE INSTALLED INTO A DIVISION 2 OR ZONE 2/22 ENVIRONMENT.
IN ORDER TO INSTALL THE DIVISION 2 APPROVED PANTHER PANEL MOUNT
UTILIZING THE FACTORY MUTUAL APPROVAL, METTLER TOLEDO CONTROL
DRAWING 155907R MUST BE FOLLOWED WITHOUT EXCEPTION. IN ORDER TO
INSTALL THE CATEGORY 3 PANTHER PANEL MOUNT TERMINAL USING THE KEMA
APPROVAL CERTIFICATE 02atex1163x ALL LOCAL REGULATIONS MUST BE
FOLLOWED WITHOUT EXCEPTION. FAILURE TO DO SO COULD RESULT IN BODILY
HARM AND/OR PROPERTY DAMAGE.
FOR EUROPEAN ZONE 2/22 HAZARDOUS AREA APPLICATIONS, THE PANTHER
CATEGORY 3 INSTALLATION GUIDE (*)16684500A AND THE KEMA TYPE
EXAMINATION CERTIFICATE MUST BE FOLLOWED WITHOUT EXCEPTION.
IF THE KEYBOARD, DISPLAY LENS OR ENCLOSURE IS DAMAGED ON A DIVISION
2 APPROVED OR CATEGORY 3 MARKED PANTHER PANEL-MOUNT TERMINAL
THAT IS BEING USED IN A DIVISION 2 OR ZONE 2/22 AREA, THE DEFECTIVE
COMPONENT MUST BE REPAIRED IMMEDIATELY. REMOVE AC POWER AND DO
NOT RE-APPLY POWER UNTIL THE DISPLAY LENS, KEYBOARD OR ENCLOSURE
HAS BEEN REPAIRED OR REPLACED BY QUALIFIED SERVICE PERSONNEL.
FAILURE TO DO SO COULD RESULT IN BODILY HARM AND/OR PROPERTY
DAMAGE.
THE PANTHER TERMINAL HAS A FACTORY MUTUAL TEMPERATURE RATING OF
T3A (180˚ C) AND A KEMA TEMPERATURE RATING OF T4 (135˚ C). IT MUST NOT
BE USED IN AREAS WHERE THE AUTO IGNITION TEMPERATURE OF THE
HAZARDOUS MATERIAL IS BELOW THIS RATING.
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING