Sirona Dental Systems GmbH
5.2
Positioning the patient/X-ray tube unit
60 81 967 D 3350
D 3350.201.02.08.02
Adjust the X-ray tube unit
Have the patient sit down on the chair.
Right-angle technique (with X-ray field limitation)
Position the film or X-ray sensor using a holding system.
When employing the right-angle technique, the X-ray
sensor should be used only with holders designed for
use with our sensor holder.
Always follow the Operating Instructions for intraoral
radiography supplied with SIDEXIS.
45° angle technique (without X-ray field limitation)
The film or X-ray sensor must be held by the patient.
i
NOTE
The sensor should be placed in a new hygienic pro-
tective cover prior to each application on a new pa-
tient!
The protective sanitary covering is not shown in the illus-
trations for the sake of clarity.
Angle of inclination
Exposure tube for the occlusal plane
Supramaxillary teeth
Molars
Bicuspids and canine teeth
Front teeth
Bite-block exposure
Bite-block exposure
Front teeth
Bicuspids and canine teeth
Molars
Submaxillary teeth
5 Operating
35°
45°
55°
80°
-90°
-20°
-10°
-5°
15