6 Operation
6.2
Positioning the Patient
A
9
,,,,,,,
10
P I
20
c
s
66
I0
• Have the patient remove all metallic objects, such as
glasses and jewelry, from the head and neck regions.
Have him take out removable dentures.
• Physical constitution, clothing, bandages etc. must not
interfere with the functioning of the unit! Perform a test
run with the T button, (see page 19)
• Insert bite block / contact segment or head positioner,
see page 18
For selection see chapter "Exposure Programs".
4
• Insert loaded film cassette (9) in the cassette carriage
until the lock (10) engages.
The cassette locks A must be placed up!
• After positioning the patient swing in the cassette. The
LED over the R key at the Multitimer switches off to let
you know that the unit is ready for the exposure.
Help Message
Should the LED over the R key go on blinking, call up the
4
help message H3 to look for the reason.
List of Help Messages see page 29.