CARE AND CLEANING

This chapter describes general care and cleaning of the TheraKair Visio MRS. Both scheduled and
unscheduled maintenance are described along with preventative maintenance. Information on transport
and storage of the system while not in use is also detailed.
The following are the ArjoHuntleigh recommended cleaning procedures for the TheraKair Visio MRS and
periodic maintenance for the therapy unit while in use.

Daily Care and Cleaning of TheraKair Visio MRS While in Use

i
The patient need not be removed from the TheraKair Visio MRS when performing daily
cleaning procedures.

Mattress

1. Press the InstaFlate function button on the Home screen to access the InstaFlate function screen.
2. Press the ON button on the InstaFlate function screen to activate InstaFlate function.
3. Verify the side rails on both sides of the bed are locked in the full upright position.
4. Gently roll the patient to one side of the mattress.
5. Wipe the cover sheet as required with soap and water or approved hospital germicide, according to
individual hospital policy.
6. Extensively soiled cover sheet or mattress should be replaced and laundered. Refer to the Weekly
Care and Cleaning section of this chapter for cleaning procedures.
7. Return the patient to the center of the mattress.
8. If InstaFlate function is still active, press the OFF button on the InstaFlate function screen and exit
the screen to resume therapy at previous settings.

Therapy Unit Screen

The Therapy Unit screen should be cleaned daily, or more often as required.
The Control screen display should be locked before cleaning.
Pressing
on the Home screen, and holding for three seconds will lock / unlock the control screen display.
The screen may be wiped with a cloth dampened in water and mild detergent.
i
Use only non-abrasive cleaning products on the therapy unit and control screen.