WARNING!
Additional equipment
Additional equipment connected to medical electrical devices must be demon-
strated to be compliant with their respective IEC or ISO standards (IEC 60601-1,
IEC 60950 or IEC 62368 for data processing equipment). Furthermore, all configura-
tions must comply with the normative requirements for medical systems (see sec-
tion 16 of the last valid edition of IEC 60601-1). Anyone who connects additional
devices to medical electrical equipment is a system configurator and as such is re-
sponsible for the system's compliance with the normative requirements for sys-
tems. Please contact the technical service if you have additional questions.
5.3
Fluid Monitoring Unit Set-up
CAUTION!
Combination of AQL-100PBS with AQL-100CBS
The irrigation pump unit AQL-100PBS may only be used with the fluid monitoring
unit AQL-100CBS as only this combination is approved to bear the MET NRTL mark.
The fluid monitoring unit (cart/scale) consists of a weighing unit for fluid bags (1),
a weighing unit for fluid containers (5), and a roller wheel base (8).
1.
Remove the cart/scale from the cardboard shipping box.
2.
Remove the pump and the power cords from the first cardboard box.
Loosen the handwheel (6) (Fig. Scale and pump connection [} 24]) and pull
3.
bag scale upwards to stop. The screw (7) (Fig. Scale and pump connection
[} 24]) must be inserted into the provided opening. Secure the bag scale with
the handwheel.
4.
Depending on the type of container used, attach the container rings (included
in the second box) to the upper (6) or lower (7) container holders (Fig. Fluid
monitoring Unit (cart with scale) [} 23]).
Fig. 5–3 Fluid monitoring Unit (cart
with scale)
(1)
(1)
(2) Fluid bags
(3) Bag deflector
(4) Container scale cable/connector
(2)
(5) Container scale
(6) Holder for upper containers
(3)
(7) Holder for lower container
(8) Roller wheel base
(4)
(9) Container
(5)
(6)
(7)
(8)
(9)
System Operation
Bag scale
(Serres, Medela)
(Abbott, Bemis®, Medi-Vac®,
DeRoyal®)
23 / 378