Electromagnetic Compatibility
en
Electromagnetic
ference immunity tests
Electrostatic discharges
(ESD) IEC 61000-4-2
Electrical
fast
ents/bursts according to
IEC 61000-4-4
Surges
according
IEC 61000-4-5
Blackouts,
brownouts,
and fluctuations of the
power supply according
to IEC 61000-4-11
Supply
(50/60 Hz)
field IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Transients RF
IEC 61000-4-3
52 / 378
12.3 Guidelines and Manufacturer's Statement - Electromagnetic
Interference Immunity
The device Aquilex® Fluid Control System is intended for use in an electromagnetic
environment as described below. The user/operator of the device Aquilex® Fluid
Control System should make sure the device is operated within such an environ-
ment.
inter-
IEC 60601 test level
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV for power lines
transi-
Repetition
quency 100 kHz
± 1 kV line(s) to line(s)
to
± 2 kV line(s) to earth
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycles 0 % UT; 250/300 cycles
frequency
magnetic
30 A/m
3 Vrms
0.15 MHz to 80 MHz
6 Vrms in ISM bands
between 0.15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m 0.15 MHz to
80 MHz
80 % AM by 1 kHz
Compliance levels
± 8 kV contact
± 2, ± 4, ± 8, ± 15 kV air
± 2 kV for power lines
fre-
Repetition
fre-
quency 100 kHz
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle and
70 % UT; 25/30 cycles
Single phase: at 0°
30 A/m
3 Vrms
0.15 MHz to 80 MHz
6 Vrms in ISM bands
between 0.15 MHz and
80 MHz
80 % AM at 1 kHz
3 V/m 0.15 MHz to
80 MHz
80 % AM by 1 kHz
Electromagnetic environment guidelines
Floors should be made from wood or concrete or
covered with ceramic tiles. If the floor covering
consists of synthetic material, the relative humid-
ity should be at least 30 %.
The quality of the supply voltage should be the
same as the voltage of a typical business or hos-
pital environment.
The quality of the supply voltage should be the
same as the voltage of a typical business or hos-
pital environment.
Magnetic fields of the mains power frequency
should comply with the typical values of business
and hospital environments.