Descargar Imprimir esta página

Changing Bags During Surgery - Hologic Aquilex Instrucciones De Uso

Ocultar thumbs Ver también para Aquilex:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 190
System Operation
en
System operation
Completing system operation
Bag change during surgery
36 / 378
NOTE!
To stop the pump
The pump continues to operate after automatic lumen calibration is complete. The
pump should be stopped by closing hysteroscope inflow stopcock.
CAUTION!
New lumen calibration
A new lumen calibration must be carried out with each change of the hysteroscope
during surgery (see Section Changing Instrument during Surgery [} 37]).
• Open stopcock and guide the hysteroscope with fluid flowing into the uterus.
• Adjust intrauterine pressure setting as necessary to obtain adequate disten-
sion and visualization.
• After the medical procedure, if system operation is complete close the hystero-
scope inflow stopcock.
• Wait until the entire fluid volume from the drape and the tube set has been
collected in both containers.
• Press the Pause/Resume button.
• Note the deficit volume indicated on the deficit display. This is the total fluid
volume that was absorbed by the patient.
WARNING!
System error
Do not use the Aquilex® System if a defect is suspected or detected during the
function test. This also applies to any obvious defects, especially defects on the
power connector or plug and power cord.
WARNING!
Failure of scale connection
If the message "Check Scale Connection" appears, the deficit must be calculated
manually. The pump keeps displaying the last known deficit value determined prior
to the failure of the scale connection.
NOTE!
Change the containers and bags during surgery
It is possible to change the containers and bags during surgery without losing the
previously measured deficit.

5.13 Changing Bags during Surgery

The device system detects automatically the replacement of a bag. Brief fluctu-
ations in the deficit calculation (< 10 s) may occur when a bag is replaced. The re-
placement of a bag is indicated with the message Bag Change, Please Proceed.
• Close the clamp of the empty bag.
• Hang a new fluid bag to a hook of the bag scale.
• Reconnect the new fluid bag with the inflow tube set.
WARNING!
Touching the bags and the hooks
Touching the bags and the hooks as well as vibrations of the balancing system
should be avoided during surgery to prevent false detection of the bag change and
not negatively affect the accuracy of the deficit calculation.