System Operation
Fig. 5–6  High vacuum tube
en
(6) Hygiene filter
(7)
Vacuum tube (green
connectors)
(8) Connection for high vacuum
(green)
(9) Container
(10) Tissue trap
(11) Adjustment controller
(12) Connecting tube
Fig. 5–7  Front of Device
28 / 378
(11)
(6)
onl y
My oS
for
ure
®
(7)
(8)
5.4
Turning On the Aquilex® System
1.
Press the ON/OFF switch. The displays and indicators light up and system turns on.
2.
The system now performs a device self-test. The device self-test is used by the
system to check whether the container and the bag scale are connected,
among others. If the bag scale is not connected, the message System OK. Bag
Scale Not Connected is displayed. In this case, please check the appropriate
data connection of the scales with the pump according to Chapter Fluid Monit-
oring Unit Set-up [} 23].
3.
If a tube set is in the inflow tube holder when the pump is switched on, the
pump display (Fig. Front of irrigation pump unit [} 22] (1)) shows the message
Remove Tube Set. The device self-test resumes once the tube set is removed
from the roller wheel.
If the device self-test is unsuccessful, the corresponding error messages are
displayed (see Chapter Error and Warning Messages [} 48]).
The system has successfully completed the device self-test when a single audible
beep is heard. The message System OK is displayed for 5 seconds followed by the
message Insert Tube Set.