In order to cut clamped and sealed tissue, use the blade activation lever 6.
Risk to patient and damage to the instrument due to clamping of metallic objects!
Do not seal or cut over clips or clamps.
WARNING
If an error report appears:
– Do not cut clamped tissue.
– Loosen the lock of the handle and open the jaw part.
– When sealing quality is sufficient, the sealed tissue can be cut using surgical scissors or the instrument.
– When sealing quality is insufficient, the sealing process must be restarted. If the additional HF cycle leads to
a normal seal, the tissue can be cut, and the instrument can be opened and removed.
Do not leave the jaw actuation lever in the locked position when the instrument is not in use.
Single-use products
Infection hazard for patients and/or users and impairment of product functional-
ity due to reuse. Risk of injury, illness or death due to contamination and/or
impaired functionality of the product!
Do not reprocess the product!
WARNING
Storage
Store sterile packed single-use products dust-protected in a dry, dark and temperature-controlled room.
Troubleshooting
Errors, faults and warnings are displayed on the Lektrafuse HF generator as follows:
An individual sound or a clear sound sequence and interruption of the sound sequence of HF operation
The yellow regrasp error light or the red error light on the front side of the Lektrafuse HF generator lights up
Notification in the display on the front side of the Lektrafuse HF generator
For details on errors, faults, warnings and remedying such, see the instructions for use of the Lektrafuse HF gen-
erator.

Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and for-
feiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95-1602
Fax:
+49 (7461) 16-5621
E-Mail:
ats@aesculap.de
Other service addresses can be obtained from the address indicated above.
Disposal
Adhere to national regulations when disposing of or recycling the product, its components and its packaging.