Troubleshooting list

Problem
Image cloudy
Image too dark, too little
illumination
9.1

Repairs

For repairs, contact the manufacturer or authorized repair center. Autho-
rized repair centers can be inquired about from the manufacturer.
For a fast processing of your service requests, send in the product indicat-
ing:
Article number (REF)
Serial number (SN)
Detailed description of defects
Note
If a component of the visualization system is replaced (e. g. service,
upgrade), the system start-up process must be performed again. It is also
recommended to perform a safety inspection after any maintenance or
replacement of a device.
Possible cause
Glass surfaces soiled
Stubborn residue on the glass surfaces
Glass surfaces soiled
Light guide dirty, defective
Excessive chloride concentration
Heavy metal ions and/or silicates, increased content of
iron, copper, manganese in water
Too high concentration of mineral substances (e.g. cal-
cium) or organic substances
Contaminated cleaning/disinfecting solutions, too fre-
quent use
Extraneous rust, (e.g. resulting from previously dam-
aged or non-corrosion-resistant instruments being
reprocessed at the same time)
Contact corrosion
Remedy
Clean the glass surfaces according to the Cleaning and Disin-
fection section.
Remove residues according to the Cleaning and Disinfection
section; check water quality.
Clean the glass surfaces according to the Cleaning and Disin-
fection section.
Check light guide (e. g. illuminate onto a white surface).
Check water quality.
Check water quality; if applicable, only use deionized (fully
desalinated) water.
The cleaning and disinfecting solutions should be replaced
regularly.
Check supply systems; in case of joint reprocessing, check
material compatibility and for existing damage, and avoid
mutual contact.
Avoid contact with other components.

10. Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
Malfunction of the product due to damage in
transport!
CAUTION
Select suitable and safe packaging (ideally the
original packaging).
Keep the original packaging for possible returns
in case of service.
Package the product in such a way that the
packaging does not become contaminated.
Note
Service and repair work must only be carried out by authorized trained per-
sonnel. For service and repairs, please contact your national
B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in
loss of warranty rights and applicable licenses.