57
ENGLISH
8.2 Technical parameter over-limit prompt
When power is about to exhaust and prompt is ON, the prompt will occur. This prompt can not be canceled unless being
closed or the power replaced.
Chapter 9
THE USAGE METHOD OF SPHYGMOMANOMETER
9.1 Accurate Measurement Way
The limb chosen for taking the measurement should be placed at the same level as the animal's heart. If this is not possible
you should apply the following corrections to the measured values: If the cuff is placed higher than the heart level, add 0.75
mmHg for each inch of difference. If it is placed lower than the heart level, deduct 0.75 mmHg for each inch of difference.
Advice
Try to measure your animal's blood pressure at the same time each day with the same limbs and the same pose
for consistency.
Do not touch Electronic Sphygmomanometer, cuff and windpipe during measurement.
Measurements should be taken in a quiet place and the body relax.
Remain still 4~5 minutes before measurement.
Do not bark or move during the measurement,Relax the body, do not let the muscle activity.
Wait 4~5 minutes between measurements.
Do not use precision instruments near the Sphygmomanometer.
Please use the sphygmomanometer at an environment of suitable temperature and humidity(refer to Chapter 18, otherwise
it will cause measurement error.
Warning
When repeatedly measuring, the accurate blood pressure value may not be measured due to congestion in the limbs.
Please measure after the blood flow is smooth.