Chapter 12 Spo 2 Measurement Method - Gima CONTEC08A-VET Manual Del Usuario

Esfigmomanómetro veterinaria electrónico
Ocultar thumbs Ver también para CONTEC08A-VET:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Chapter 12
SpO
MEASUREMENT METHOD
2
(This chapter is only suitable for European Union market)
1) Put the animal tongue enough into the animal nip of probe as the following figure.
SpO
Place
2 probe
2) Plug the connector of the SpO
switch to SpO
interface. This operation brings no affection to other functions.
2
Note
SpO
display range: 0% ~ 100%, PR display range: 30 bpm (beats/min) ~ 250 bpm (beats/min)
2
If the SpO
works abnormally, after connecting the SpO
2
face or no data displayed under the SpO
Measurement Limitation
During operation, the accuracy of SpO
• High-frequency electromagnetic interference such as interference from electrosurgical apparatus connected to the sys-
tem.
• Intravenous dyestuff.
ENGLISH
probe cable into the USB socket in the lower right of the device. The main interface will
2
interface.
2
readings can be affected by:
2
probe to the device, the device will not switch to the SpO
2
66
inter-
2

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

80550

Tabla de contenido