Riester ri-focus LED Instrucciones Para El Uso página 27

Lámpas frontales y espejos frontales
Ocultar thumbs Ver también para ri-focus LED:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 44
The ri-focus LED headlights and clar N LED head mirrors are intended for use in the electromagnetic environment specified below. The customer or user of the
headlight should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
Con: ±8 kV
(ESD)
Air: ± 2, 4, 8,
IEC 61000-4-2
15 kV
Rapid transients electri-
Not applicable
cal disturbances/bursts
IEC 61000-4-4
Impulse voltage
Not applicable
IEC 61000-4-5
Voltage dips, short
Not applicable
interruptions and voltage
variations according to
IEC 61000-4-11
Magnetic field with
30A/m
energy-efficient rated
frequencies
IEC 61000-4-8
Note: UT is the AC source. Mains voltage before the application of the test level.
27
Guidelines and manufacturer's declaration – electromagnetic immunity
Compliance
Con: ±8 kV
Air: ± 2, 4, 8,
15 kV
Not applicable
Not applicable
Not applicable
30A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If the floor is covered
with synthetic material, the relative humidity must be at least 30%.
The quality of the supply voltage should be that of a typical commercial
or hospital environment.
The quality of the supply voltage should be that of a typical commercial
or hospital environment.
The quality of the supply voltage should be that of a typical com-
mercial or hospital environment.
Magnet fields at mains frequency should be at a level typical for the
location of a typical commercial hospital environment.

Publicidad

Tabla de contenido
loading