Fault/error code
019-272XX-XXX

10.3 Data transmission

Fault/error code
The "Date of birth" dialog
window appears after
pressing the
key
Autoclear function: Entering
"0 seconds" is not accepted;
instead, the factory setting
(300 seconds) is suggested
Workflow LED does not
light up
Workflow LED lights up red
02-288XX-XXX
052-280XX-XXX
420-280XX-XXX
104 •
seca 651/650
Cause
Ambient temperature too high or too
low.
Cause
Patient ID not scanned
ID scanning not set up
Patient does not yet have a file in the
EMR System
Device is connected to the
software
seca analytics 125
Implausible input; switching off the
function is not intended in the factory
No network connection
WiFi function deactivated
Workflow LED is defective
ID was
found in the EMR System or
not
the seca software
Device has
saved measuring
not
results to the clipboard.
The measuring results were
submitted to an EMR System or
seca software.
No data transmission possible
Internal device communication error
No connection to the server
• Observe ambient conditions for operation,
transport, and storage ➔
technical data, page 109
• Press
• Scan patient ID
Check workflow settings for the device in
the
seca connect 103
• Enter date of birth and press
again
• Create patient file in the EMR System and
assign measurement
• Enter date of birth
• Press
• Accept factory setting
• Manually enter value between 1 and 3600
seconds
Set up network connection ➔
network functions, page 95
Activate WiFi function ➔
deactivating the WiFi function, page 97
Inform seca Service
Create the ID in the EMR System or the
seca software
Repeat measurement
not
• Repeat measurement
• Check WiFi connection
Check network and workflow settings:
• Check network settings ➔
network functions, page 95
• Check workflow settings:
seca connect 103
use
• Disconnect the plug-in power supply unit
from the socket
• Wait approx. 1 minute
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Inform seca Service
• Check LAN cable
• Check network settings ➔
network functions, page 95
Remedy
General
Remedy
key
software
key
key
Setting up
Activating/
Setting up
system instructions for
Setting up