Fault
Multifunctional display
does not react when keys
are pressed
Multifunctional display
remains dark after
pressing
10.2 Measuring weight
Fault/error code
Displayed weight value is
implausible
Autohold
function cannot
be deactivated
001-272XX-XXX
to
008-272XX-XXX
020-272XX-XXX
to
023-272XX-XXX
010-272XX-XXX
013-272XX-XXX
Cause
Device is in an undefined state
following implausible inputs.
With device combinations with a
standing aid and/or measuring rod:
Additional display connected
Device is in an undefined state
following implausible inputs.
Multifunctional display defective
Cause
Weighing electronics use outdated
zero point
Weighing electronics defective
Device is connected to a network
(intended behavior): Autohold function
is activated automatically.
Load cell or weight calculation module
defective
One corner of the scale has been
loaded excessively.
The scale has been switched on with
too high a load.
The scale was caused to oscillate and
was unable to determine the zero point.
Remedy
• Disconnect the plug-in power supply unit
from the socket
• Wait approx. 1 minute
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Remove the second display
• Disconnect the plug-in power supply unit
from the socket
• Wait approx. 1 minute
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Disconnect the plug-in power supply unit
from the socket
• Wait approx. 1 minute
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
Inform seca Service
Remedy
• Remove the weight from the weighing
platform
• Press Weight display field
• Press
key: Device restart
• Wait until main screen is displayed again
Inform seca Service
If necessary, disconnect device from
network
Inform seca Service
• Distribute weight evenly
• Disconnect the plug-in power supply unit
from the socket
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Inform seca Service
• Remove the weight from the weighing
platform
• Disconnect the plug-in power supply unit
from the socket
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
• Disconnect the plug-in power supply unit
from the socket
• Plug the plug-in power supply unit into the
socket; the device and the multifunctional
display switch on automatically
Troubleshooting • 103
English