SERIAL COMMUNICATION AND PRINTER SETUP CONTINUED
Status Byte B Bit Definitions
STATUS BITS FUNCTION
Bit 0
Gross = 0, Net = 1
Bit 1
Sign, Positive = 0, Negative = 1
Bit 2
Out of Range = 1 (Over Capacity or Under Zero)
Bit 3
Motion = 1, Stable = 0
Bit 4
lb = 0, kg = 1, (see also Status Byte C, bits 0, 1, 2)
Bit 5
Always = 1
Bit 6
Zero Not Captured after power-up
The Indicator supports both MT-SICS and OHAUS commands. Commands listed in the following tables will be
acknowledged by the indicator. To use the MT-SICS commands, send the command PSI. To return to the OHAUS
commands, send the command POH. SICS commands can also be active in the menu setup.
OHAUS Commands
COMMAND
FUNCTION
IP
Immediate Print of displayed weight (stable or unstable).
P
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
xS
0S: Turn off "Stable Only" menu item and allow unstable print. 1S: Turn on "Stable Only" menu item and only print stable print.
xP
Interval Print x = Print Interval (1-50,000 sec), 0P turns auto print OFF.
Z
Same as pressing Zero Key.
T
Same as pressing Tare Key.
Download Tare value in grams (positive values only).
xT
Sending 0T clears tare (if allowed).
PU
Print current unit: g, kg, lb, oz, lb:oz
xU
Set scale to unit x: 1=kg, 2=lb, 3=g, 4=oz, 5=lb:oz
xM
Set scale to mode X. 1=Weighing, 2=Counting, 3=Check, 4=Percent, 5=Dynamic. M will scroll to next enabled mode.
PSN
Print Serial Number.
CU xxx
Set Under Limit (only in Check mode) where 'xxx' is the value under current unit.
CO xxx
Set Over Limit (only in Check mode) where 'xxx' is the value under current unit.
x#
Set Counting APW (x) in grams. (only in Counting or Check Counting mode, must have APW stored).
P#
Print Counting or Check Counting mode APW.
x%
Set Percent mode reference weight (x) in grams (must have a weight stored).
P%
Print Percent mode reference weight.
PV
Version: print name, software revision and LFT ON (if LFT is set ON).
H x "text"
Enter String content, x = String number (1-10), "text" = string text up to 40 alphanumeric characters.
\EscR
Global reset to reset all menu settings to the original factory defaults.
PAGE 27 OF 101
Status Byte C Bit Definitions
BITS 2, 1 AND 0
2
1
0
0
0
0
0
0
1
0
1
0
0
1
1
1
0
0
1
0
1
1
1
1
1
1
1
BIT 3
BIT 4
BIT 5
BIT 6
Weight Description
lb or kg, selected by Status Byte B,
bit 4
grams (g)
metric tons (t)
ounces (oz)
not used
not used
tons (ton)
no units
Print Request = 1
Expand Data x 10 = 1, Normal = 0
Always = 1
Always = 0