English
Caution: The measurement probe emits an alternating magnetic field that could potentially interfere
with cardiac pacemakers!
Technical Description/Specifications
Technical description
Osstell ISQ is CE-marked according to MDD in Europe. Osstell ISQ is constructed according to,
and fulfilling the standards EN 60601-1 (Class II, Internally powered, type BF applied part. Not AP
or APG equipment. Not protected against ingress of water.), EN 60601-1-2 and UL 2601-1. The sym-
bols used, follow, to the extent possible, the European standard EN 60601-1, and the ISO-standards
9687 and 15223.
Specifications
Power, weight & size
Rated power:
8 VA, Type FW 7660M/05
Instrument size:
195 x 120 x 45 mm
Package size:
280 x 240 x 63 mm
Instrument weight: 0.4 kg
Gross weight:
1.0 kg
Environmental conditions during transport
Temperature:
-40 ºC to +70ºC
Relative humidity:
10% to 95%
Pressure:
500 hPa to 1060 hPa
Environmental conditions during use
Temperature:
+10 ºC to +40 ºC
Relative humidity:
30% to 75%
Pressure:
700 hPa to 1060 hPa
IP-class:
IP20
Accuracy
ISQ accuracy is within +/- 0,5 ISQ units for a single SmartPeg. Including variations in the attache-
ment torque and individual variations between different SmartPegs, the accuracy is +/- 2 ISQ units.
Symbols
i
Consult instructions for use
Class II Equipment
Type BF equipment
IP20
No protection against water
Do not re-use
Caution
Manufacturer
º
135 C
Sterilizable up to 135°C
Separate collection
20