Troubleshooting

Overload error
Possible cause
Head lock up holding force exceeded during CMM acceleration
Collision with workpiece or CMM
Obstruct error
Possible cause
Head / stylus configuration obstructed
Probe extension combination too long
Internal head fault
PH10 PLUS head and non-PLUS PHC10-3 or third party controller
Datum error
Possible cause
Cable / connections faulty
Head / stylus configuration obstructed
Probe extension combination too long
Internal head fault
No head movement
Possible cause
Power loss
Cable / connections faulty
Solution
Check condition of cabling from head to controller
Remove obstruction and update head using HCU1 or under CMM control
Use shorter / lighter stylus combination
Contact your Renishaw representative
Solution
Check that controller power on LED is lit
Check mains cable condition and integrity
Check controller fuse
Check emergency stop condition
Check condition of cabling from head to controller
Issued 01 2013
Page 32 of 48
Solution
Reduce mass of probe combination
Reduce acceleration
Check head for damage
Remove obstruction and repeat move
Solution
Check head for damage
Remove obstruction and repeat move
Use a shorter / lighter stylus combination
Contact your Renishaw representative
Contact your Renishaw representative