10.2 Warnings
Warnings show on the controller display. To see all of the active warnings, push menu then select
DIAGNOSTICS>TU5x00 sc>WARNING LIST.
Warning
CLEANING
The automatic cleaning module
MODULE
does not operate correctly.
DESICCANT OLD
The desiccant cartridge is more than
2 years old.
DRYER EXHAUS'D
The desiccant cartridge life is zero.
HIGH FLOW
The flow rate is higher than the limit
(more than 1250 mL/min).
HUM PCB SC
There is humidity on the interior
electronics of the instrument.
LASER-TEMP
The laser temperature is higher than
HIGH
the limit.
LASER-TEMP
The laser temperature sensor has a
SENS
malfunction.
LOW FLOW
The flow rate is lower than the limit
(less than 75 mL/min).
NO FLOW
The flow rate is less than
10 mL/min.
NOT DRYING
The instrument cannot regulate the
internal humidity.
PUMP
The air pump for the drying circuit
has a malfunction.
SENS.DRY: FUNC
The air system of the drying system
has a malfunction.
TURB TOO HIGH
The turbidity reading is not within
the calibration range.
WIPER REPLACE
A wiper replacement is due in the
automatic cleaning module.
VIAL CLARITY
The vial or vial compartment is dirty.
Description
Solution
Make sure that the wiper head is installed correctly
and the wiper arm can move up and down.
Replace the desiccant cartridge. Refer to the
documentation supplied with the desiccant
cartridge.
Replace the desiccant cartridge. Refer to the
documentation supplied with the desiccant
cartridge.
Adjust the flow regulator as necessary. Make sure
that the flow regulator does not have a malfunction.
Contact technical support. Measurements with
limited validity are still available.
Decrease the environmental temperature of the
instrument.
Contact technical support. Measurements with
limited validity are still available.
Examine the tubing for blockages that decrease the
flow rate. Remove the blockages.
Adjust the flow regulator as necessary. Make sure
that the flow regulator does not have a malfunction.
Examine the tubing for blockages stop the flow.
Remove the blockages.
Replace the desiccant cartridge. Refer to
the desiccant cartridge
on page 28.
If the error continues, contact technical support.
Measurements with limited validity are still available.
Contact technical support. Measurements with
limited validity are still available.
Contact technical support. Measurements are still
available, but the life of the desiccant cartridge is
decreases.
Make sure that the calibration range selected is
applicable to the turbidity value of the sample.
Replace the wiper in the automatic cleaning module.
Refer to the documentation supplied with the
automatic cleaning module to replace the wiper.
Clean or dry the vial and the vial compartment.
Replace
English 29