Table of Contents
1 Additional information
2 Specifications
on page 3
3 General information
on page 5
4 Installation
on page 10
5 User navigation
on page 21
Section 1 Additional information
An expanded user manual is available on the manufacturer's website. Videos on how to install,
operate and do maintenance and troubleshooting on the TU5300 sc and the TU5400 sc turbidimeters
are available on the TU5 Series Turbidimeters playlist at
http://www.youtube.com/user/hachcompany.
Section 2 Specifications
Specifications are subject to change without notice.
Specification
Measurement method
Primary compliance
method
Enclosure
IP rating
Dimensions (W x D x H)
Weight
Power requirements
Protection class
Pollution degree
Installation category
Mounting
Operating temperature
Storage temperature
Humidity
Sensor cable length
Laser
Optical light source
Fittings
1
Water drops, puddles or runlets that will not damage the instrument may be in the inner of the
enclosure.
on page 3
Details
Nephelometry with scattered light collected at a 90-degree angle to the incident light
and 360 degrees around the sample vial
DIN EN ISO 7027
Material: ASA Luran S 777K / RAL7000, TPE RESIN Elastocon
Thermoplastic Elastomer TPS-SEBS (60 Shore) and stainless steel
Electronic compartment IP55; process head/Automatic Cleaning Module attached to
the instrument and all of the other functional units IP65
268 x 249 x 190 mm (10.6 x 9.8 x 7.5 in.)
Instrument with the process head: 2.7 kg (6.0 lb); Instrument with the optional
automatic cleaning module: 5.0 kg (11.0 lb)
12 VDC, 14 VA supplied by the SC controller
III
2
II
Indoor on a wall
0 to 50 °C (32 to 122 °F)
–40 to 60 °C (–40 to 140 °F)
5 to 95% relative humidity, non-condensing
TU5x00 sc without Automatic Cleaning Module or flow sensor: 50 m (164 ft);
TU5x00 sc with Automatic Cleaning Module: 10 m (33 ft)
Class 1 laser product: Contains a non user-serviceable class 1 laser.
850 nm, maximum 0.55 mW
Sample inlet and outlet: ¼-in. OD tubing (optional tubing adapter, ¼ in. to 6 mm)
6 Operation
on page 21
7 Calibration
on page 21
8 Verification
on page 21
9 Maintenance
on page 21
10 Troubleshooting
on page 28
®
STK40,
1
English 3