Descargar Imprimir esta página

B.Braun Dialog+ SW9 Serie Manual De Instrucciones página 3

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 13
Supplement ABPM Automatic Blood Pressure Measurement Module
YOU WILL NEED ONE COPY PER MACHINE. PLEASE COPY IF NECESSARY.
EN / AU
Supplement to Dialog Instructions for Use (IFU): Information about the option ABPM
and handling of old/new cuffs.
Read the following information and instruct all users in your organization. Add this
supplement to your Instructions for Use (IFU).
This supplement only contains information that is changed. For the complete
descriptions, refer to the existing Dialog Instructions for Use.
Symbols on ABPM Cuff
Symbol
Description
Follow operating instructions in these instructions for use when operating the ABPM
Type BF defibrillation proof applied part
Classification acc. to IEC 60601-1
Cuff is free of latex
Upper arm diameter
Cuff size: S (small), M (medium), L (large), XL(extra large)
The respective size is indicated by the rectangle around the symbol.
Marking for cuff placement
Handling of old/new cuffs with the option ABPM
In order to improve therapy outcomes and patient comfort, B. Braun offers a new series of blood pressure
cuffs for the option ABPM. To find out which module is assembled and cuff is needed, check the coupling in
your machine and compare with the pictures below. Follow the appropriate instructions.
Fig. 1 Male coupling on the machine
 Check if your Dialog contains the male coupling (Fig. 1).
 Use tubing with two female couplings (Fig. 2).
 Connect tubing with one female coupling to machine.
 Connect same tubing with the other female coupling to cuff.
 For further measurement, follow your instructions for use Dialog.
Supplement 38910463 / Rev. 2.87.30 / December 2018
Fig. 2 Tubing female/female
1

Publicidad

loading

Este manual también es adecuado para:

Dialog advancedDialog+