4.5. After Use of the Displaying Unit
Disconnect the Ambu visualization device from the displaying unit I . For disposal of
1.
the visualization device, refer to the Instruction for Use of the specific device.
2.
Turn the displaying unit OFF by pressing the power button
on the screen. Press OK to confirm shutdown. This means that the displaying unit is
powering down J . If the battery level is low, charge the displaying unit (see section 4.2).
3.
If needed, dismount the displaying unit from the bracket. Use two hands to hold the
displaying unit, while pressing the two grey release buttons on the bracket behind the
displaying unit. Pull the displaying unit towards yourself K .
4.
Clean and disinfect the displaying unit (see section 5).
5.
If the power button
the battery level is below 20%. In this case, charge the displaying unit (see section 4.2).
6.
Between procedures, the displaying unit must be stored in accordance with
local guidelines.
5. Cleaning and Disinfection of the Displaying Unit
The displaying unit shall be cleaned and disinfected before and after use according to good
medical practice following the below instructions.
The displaying unit is a reusable medical device. According to the Spaulding's/CDC
classification, the displaying unit is a non-critical device.
Cleaning and Disinfection procedure
The displaying unit can be cleaned and disinfected using disposable germicidal hospital
disinfection wipes designed for hard, non-porous surfaces. Compatible disinfection wipes
are based on Isopropyl alcohol 70-80%, Isopropyl alcohol/Ammonium Chloride or Isopropyl
alcohol/Ethanol.
1.
Turn OFF the displaying unit and make sure that it is unplugged from mains.
2.
Follow the cleaning instruction from the germicidal disinfection wipe manufacturer
to thoroughly clean the displaying unit.
3.
Allow the displaying unit to dry after the cleaning.
4.
Follow the disinfection instruction from the germicidal disinfection wipe
manufacturer to thoroughly disinfect the displaying unit with a new wipe.
5.
Allow the displaying unit to dry after the disinfection before use.
6.
If the cleaning and disinfection chemicals used leave a residual layer on the display
after they dry out, use an alcohol based wipe to remove the residue.
7.
If applicable, clean and disinfect the mounting bracket following the same procedure
as for the displaying unit.
NOTE: The wipes shall be moist, but not dripping to ensure no damage to internal electronics of the
displaying unit.
NOTE: If using wipes containing hypochlorite, ensure that all residue is completely removed. Wipes
containing hypochlorite may affect the screen's antireflective coatings over time. You should limit the
use of wipes containing hypochlorite to required cases only.
After cleaning and disinfection, the displaying unit must be submitted to the pre-check
procedure in section 4.2.
12
is flashing orange when the displaying unit is turned OFF,
. A notification is shown