5.4
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
For heavy soiling Erbe Elektromedizin recommends an approx.
3-percent hydrogen peroxide bath.
When using a soft plastic brush, ensure that the brush is moved
only in parallel with the coagulation surface of the BiClamp or
from the backs of the jaws in the direction towards the coagula-
tion surface. On no account brush from the coagulation surface
in the direction of the backs of the jaws. See also the following
illustration:
1. If required, place the product in an approx. 3-percent hydro-
gen peroxide bath.
2. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
3. BiClamp with MF-2 connector (21195-xxx): Ensure that the
golden contact surface on the connector is free from visible
contamination.
4. Move the movable parts back and forth (at least 5 times).
32