IV - INSTALLATION / SETUP

4.1 UNPACK THE DEVICE

Upon receiving the device, look for any damage that may have occurred during shipping.
If needed, contact your supplier.

4.2 RECOMMENDATIONS

Have all connections between your device and the power grid made by a certified dental installation technician. The power grid
to which the Genius
endodontic motor connects must comply with the applicable standards in your country.
®

4.3 INSTALLATION

Important:
Do not install the Genius
Do not insert the power cord or pedal cord in a cord cover or sleeve.
Access to the rear of the device must be easy and immediate. Disconnecting elements such as the switch and
power cable are located at the rear and must remain accessible at all times.
1-
Remove unit and its accessories from their original packaging and place them on a flat surface. Do not install near
direct or indirect sources of heat.
2-
Verify box contents, referring to Section 3.1 "PHYSICAL DESCRIPTION".
3-
Do not bend motor cord to avoid damage.
4-
All accessories are supplied non sterile.
5-
The serial number indicated on the housing rear panel should match the number on the shipping documents.
6-
Leave at least a 150 mm/6 inch space around the housing for proper cooling.
7-
Place control box on a stable, horizontal surface with a slope no greater than 5°.
8-
Verify that power switch is in O (off) position.
9-
Connect power cord to unit base.
10- Insert plug complying with transformer standards applicable in your country, and connect it all to a power outlet.
11- Insert the two gray silicon pads onto the two metal supports.
12- Connect control pedal to pedal connector.
13- Place pedal where it is easily accessible to the foot.
14- Connect endodontic motor and cord to connector.
15- Place motor on support.

4.4 FIXING THE MEDICAL DEVICE TO A NON-REMOVABLE SUPPORT

Your medical device is not designed to be moved. To avoid accidentally dropping the device, we recommend that you fix it
in a precise place in your treatment room, using the mounting kit supplied in the packaging box, to ensure that it cannot be
dismantled or moved without needing a tool.
V - SETTINGS / PARAMETERS / INTERFACE

5.1 BEFORE USE

Before powering unit on, verify that ALL COMPONENTS have been installed.
The first time it is turned on, the device displays the message "Rotating Mode", along with values set by the manufacturer. (Fig.
5)
Use the "MODE" button to switch from traditional rotation ("Rotating") to forward-reverse rotation ("F+R").

5.2 PARAMETERS

The Genius
endodontic motor stores each program parameters upon their setting by user. They are kept in memory, even if the
®
unit is powered off. It is possible to configure and store the following parameters in each program: speed, torque, contra-angle,
reduction ratio, autoreverse type, and angular values in forward and reverse rotation. To restore factory settings (software
reset), press "Mode" button while powering the device on (Fig. 5).

5.3 TORQUE CALIBRATION

To calibrate the torque, press the "Calib." button on the front panel, with the motor running. This function verifies the
handpiece mechanical inertia in order to compensate that value. This operation takes a few seconds, during which the motor
runs at various speeds (Fig. 6).
Calibration duration: Approximately 8 seconds. Do not perform any operation during calibration. Calibration should be
performed before each endodontic session and whenever the contra-angle is replaced or lubricated.
The file should not be in use during calibration.
Calibration should be performed before each treatment.
6
Operating Manual ● Genius
®
endodontic motor near or on another device.
®
Endodontic Motor ● J28629 ● V4 ● (16) ● 10/2016 ● NO69US010F