Accu-Chek Performa Nano Instrucciones De Uso página 48

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

46
4. Put the meter on a flat surface,
such as a table.
5. Remove the bottle cap. Wipe
the tip of the bottle with a
tissue.
6. Squeeze the bottle until a tiny
drop forms at the tip. Touch the
drop to the front edge of the
yellow window of the test strip.
Do not put control solution on
top of the test strip. When you
see
flash, there is enough
control solution in the test strip.
OR
(for example)
7. Wipe the tip of the bottle with a
tissue. Cap the bottle tightly.
8. The control result appears on
the display, along with the
control bottle symbol and
flashing L. Do not remove the
test strip yet. Press
once to
mark the control result as a
Level 1. Press
a second
time to mark the control result
as a Level 2.

Publicidad

Tabla de contenido
loading

Tabla de contenido