Technical Information
Use

Startup

If the forepump and the vent device are not operated by the
controller, close the vent valve and switch on the forepump.
Apply mains to the I/O connector. Voltage and frequency output will
be at the maximum level; then the frequency will decrease to a value
proportional to the pump rotational speed (3 KRPM if the pump is
completely stopped) and then it will accelerate until the normal
condition has been reached.
The controller with the Soft Start mode allows the pump to ramp-up
to Normal speed slowly with a minimum ramp-up time of 250
seconds and a maximum of about 45 minutes. If the Soft Start mode
is deselected, the ramp-up will be done within
180 seconds. Plug the controller power cable into a suitable power
source.

Monitor-Mode

The following instructions are applicable to Turbo-V 250 Controller model 969-
NOTE
9504 only when the optional "hand held terminal" is installed.
S
214/240
After the main is applied to the I/O connector, the pump starts
and the display shows:
O
F
T
[
-
-
Turbo-V 250 Controller User Manual / 87-900-856-01 (E)
S
T
A
R
T
-
-
-
-
-
O
N
-
-
-
]