Quick Start programming
1. Probe Model [see Probe Nameplate (7MLxxxx-x) (7xX) for details]:
_________________________
2. Probe Mounting Type NPT, BSP , or flanged [use part number to establish Probe
Mounting Type (NPT and BSP differ in insertion length)]: _________________________
3. Measurement Type (level only, volume, interface level, or interface level and
volume):_________________________
4. Level Units (inches, centimeters, feet, meters): _________________________
5. Probe Length [listed on model information at top of probe
(Note Y01 = xxx cm)]: ____________________
6. Level Offset (desired level reading when liquid is at the end of the probe): __________
7. Dielectric (dielectric constant range of process medium, upper layer dielectric for
interface applications): _________________________
8. Loop Control (Is output current to be controlled by level or volume?): ______________
9. Set 4.0 mA (What is the reference point for the 4.0 mA value?): __________________
10. Set 20 mA (What is the reference point for the 20.0 mA value?): _________________
Quick Setup: steps 1 to 10
1. Select Probe model (Default: 7xA-x)
Select from:
7xA-x (7ML1301-1), 7xB-x (7ML1302-1), 7xD-x (7ML1301-2), 7xF-x (7ML1303-1), 7xF-E
(7ML1303-1D), 7xF-F (7ML1303-1E), 7xF-4 (7ML1303-1J), 7xF-P (7ML1303-1H), 7xJ-x
Values
(7ML1303-2), 7xK-x (7ML1305-1), 7xP-x (7ML1301-3), 7xR-x (7ML1301-4), 7xS-x (7ML1301-5),
7xT-x (7ML1301-6), 7x1-x (7ML1304-1), 7x2-x (7ML1304-2), 7x5-x (7ML1302-2), 7x7-x
(7ML1302-3)
2. Select Probe mounting type (Default: NPT)
NPT
Values
BSP
Flange
Page EN-8
2
10
1
7
9
8
SITRANS LG200 – QUICK START MANUAL
4
5
6
7ML19985XG81