d. Remove both fastening screws for the trim plate (Fig. A 11).
e. Remove the trim plate.
f.
Loosen the dummy plug (Fig. A 6).
g. Mount the cable gland. Follow the instructions in the documentation of
the cable gland manufacturer.
3. Connect the connecting cable:
a. Mount the 2-wire connecting cable in the cable gland, cut to length and
strip (approx. 80 mm).
b. Connect both wires of the connecting cable to connection terminals. Pay
attention to the polarity.
4. Complete the installation:
a. Insert the trim plate.
b. Fasten the trim plate with both fastening screws.
c. Put the protective cap (Fig. A 3) on the battery contacts in the housing.
5. Close the gas detector:
a. Place the housing cover in the final position (Dräger logo horizontal).
b. Turn the housing cover under slight pressure approx. 20°
counterclockwise until the housing cover rests on the sealing ring on the
housing.
c. Press the housing cover forcefully and evenly onto the sealing ring.
d. Slowly turn the housing cover approx. 20° clockwise to the stop in the
end position.
e. Screw in the locking screw (Fig. A 14) until the housing cover is
prevented from being screwed open.
6. Configure the gas detector with PolySoft for operation with wired power
supply.
6.2

Set up battery operation

The transport, insertion and replacement of the battery box in potentially
explosive gas atmospheres is permitted.
In a dust explosion hazardous area, the short-term transport of the battery box
is permissible. Inserting or changing the battery box in a potentially explosive
dust atmosphere is only permitted with declassification.
Instructions for use
|
Dräger Polytron
®
6100 EC WL, Dräger Polytron
Replacing the battery pack is also explained in PolySoft .
6.2.1

Removing the battery box

WARNING
Risk of explosion!
If the gas detector is opened in a potentially explosive dust atmosphere, an
explosion may occur.
► Do not open the gas detector in dust explosion-hazard areas. Alternatively,
declassify the dust explosion-hazard area beforehand.
Remove the battery box:
1. Open the gas detector:
a. Loosen the set screw (Fig. A 14).
b. Turn the housing cover (Fig. A 1) approx. 20° counterclockwise and
remove it.
2. Pull out the battery box (Fig. A 2).
6.2.2

Installing the battery pack

WARNING
Risk of explosion!
Opening the battery box in a hazardous area can cause an explosion.
► Do not open the battery box in an explosion-hazard area. Alternatively,
declassify the explosion-hazard area beforehand.
NOTICE
Incorrect calculation of the remaining charge!
► When inserting or changing the battery pack, only use new battery packs.
Exception: If the currently used battery pack is removed, e.g., during
maintenance, it may be used again.
Prerequisite:
Battery box has been removed.
Battery box is located outside the explosion-proof area
®
Repeater ISA
|
en
Electrical installation
39