Troubleshooting
Bit
DIAG_STATUS
16
– (Bit not used)
15
– (Bit not used)
14
OTHER_FAULTS
13
– (Bit not used)
12
– (Bit not used)
11
– (Bit not used)
10
– (Bit not used)
9
– (Bit not used)
8
– (Bit not used)
7
– (Bit not used)
6
SENSOR_WARMUP
5
SENSOR_CAL_INTERVAL_EXPIRED
4
SENSOR_VITALITY_POOR
3
NO_NETWORK_CONNECTION
2
– (Bit not used)
1
– (Bit not used)
0
– (Bit not used)
48
General status
Cause
F
Serious device fault, various causes.
O
Warm-up phase 1 active.
M
Calibration interval expired.
M
Vitality of the sensor is low.
F
Incorrectly configured
ISA100 Wireless
Instructions for use
|
Dräger Polytron
Remedy
Have the gas detector checked by Dräger-
Service.
Wait for warm-up phase 1 to finish.
Calibrate the gas detector.
Provide for sensor change promptly.
Check information provision.
network.
®