Hach TL2360 Manual Del Usuario página 3

Ocultar thumbs Ver también para TL2360:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 89
Table of contents
Specifications
Specifications are subject to change without notice.
Specification
Measurement method
Regulatory
Dimensions (W x D x H) 39.5 x 30.5 x 15.3 cm (15.6 x 12.0 x 6.02 in.)
Weight
Enclosure
Protection Class
Pollution degree
Installation category
Power requirements
Operating temperature
Storage temperature
Humidity
Display
Light source
Measurement units
Range
Nephelometric
Meets ISO 7027, DIN EN 27027, DIN 38404 and NFT 9033
ASTM D7315 - Standard Test Method for Determination of Turbidity Above 1 Turbidity
Unit (TU) in Static Mode
ASTM D6655 - Standard Test Method for Determination of Turbidity Below 5 NTU in
Static Mode
2.9 kg (6.4 lb)
IP30; indoor use only
External power supply: Protection Class I; instrument: Protection Class II
2
External power supply: Category II; instrument: Category I
Instrument: 12 VDC, 3.4 A; power supply: 100-240 VAC, 50/60 Hz
0 to 40 °C (32 to 104 °F)
-20 to 60 °C (-4 to 140 °F)
5 to 95% relative humidity, non-condensing
17.8 mm (7 in.) color touch screen
Light-emitting diode (LED) at 860 ± 30 nm
FNU, NTU, EBC, Abs (absorbance), %T (% transmittance) and mg/L (degree)
FNU (Ratio on): 0-1000
NTU (Ratio on): 0-10,000 auto decimal
EBC (Ratio on): 0-2450 auto decimal
Absorbance (auto range): 0-2.00
Transmittance (%): 1.0-100
Degree: 0-100 mg/L
Startup
on page 10
Operation
on page 11
Maintenance
on page 19
English 3

Publicidad

Capítulos

Tabla de contenido
loading