Descargar Imprimir esta página

Cleaning, Disinfection, Sterilization; Safety Instructions For Processing; Processing Limitation; Dismantling - Erbe 20132-106 Manual Del Usuario

Apc-aplicadores

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 72
OBJ_DOKU-189911-006.fm Seite 30 Donnerstag, 17. Juni 2021 9:31 09
8

Cleaning, disinfection, sterilization

8.1

Safety instructions for processing

Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
Suitable for a maximum of 20 ultrasonic cycles.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of validated
procedures".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
8.2

Processing limitation

Frequent processing can damage this product. Observe the safe-
ty instructions for checking the product. With visible damage or
functional impairment, the product may no longer be used.
8.3

Dismantling

1. Disconnect the APC applicator from the APC cable.
2. For APC-applicator 20132-106 only: Remove protection
tongue.
30

Publicidad

loading

Este manual también es adecuado para:

20132-110