Hach TL2360 Manual Básico Del Usuario página 23

Ocultar thumbs Ver también para TL2360:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 48
Message
Read/Write to USB flash drive
Cannot write to USB
memory
Cannot read from USB
memory
Restore backup
No instrument backup is
available.
Not able to restore the
backup
Security
Invalid password
Send data
Connect a receiving
device.
Add sample IDs from list
No valid data found
Not able to read sampling
date.
The instrument cannot
read the Sample ID
Problem/Error: Incorrect
date
Possible cause: The
wrong date format.
The sample ID list full.
Data has not been added.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Look for remaining space on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive to the instrument.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive. Start the instrument update
procedure again.
Enter the correct password. If the password is lost, contact technical
support.
Examine the device connections. Set the Auto Send setting to off.
Refer to
Configure the measurement settings
No sample ID file was found on the USB flash drive.
Make sure that the date and time format is dd.mm.yyyy hh:mm.
Examine the text strings. Refer to the expanded user manual on the
manufacturer's website.
Make sure that the date and time format is dd.mm.yyyy hh:mm.
Remove the sample IDs that are not used. Add a new sample ID.
Solution
on page 12.
English 23

Publicidad

Tabla de contenido
loading

Tabla de contenido