Battery charge indicator (g), temperature warning and
"ERROR" indicator in the display
Temperature
measuring tool not within
the allowable charge-
temperature range.
Battery charge-control indicator (g) and "ERROR"
indication in the display
Battery charging voltage
not correct
Measuring result not plausible
The target surface does
not reflect correctly (e.g.
water, glass).
Laser beam output 19,
reception lens 20 or
camera 18 is covered.
Wrong reference level
set.
Obstruction in path of
laser beam.
Data transmission via USB interface not possible
Micro USB cable
42
2610050173 04-18 GLM400CL.indd 42
Cause
of
the
Corrective Measure
Wait
until
temperature range is reached.
Check if the plug-in connection
has been established correctly
and if the battery charger is
operating properly. When the
unit symbol is flashing, the
battery is defective and must be
replaced by a Bosch after-sales
service.
Cover off the target surface.
Keep the laser beam output 19,
reception lens 20 and camera
18 clear.
Select reference level that
corresponds to measurement.
Laser point must be completely
on target surface.
Check the proper and tight seating
of the micro USB cable.
Check the micro USB cable for
damage.
the
charge-