If the PureWick™ Urine Collection System is dropped or tipped over spilling urine, unplug
the unit and carefully inspect for loose or damaged parts before resuming use. Contact
Customer Support at 1-888-201-1586 if any damage is observed.
The pump tubing connecting the PureWick™ Urine Collection System to the collection
canister may experience light condensation inside the tube. This is not unusual and does
not affect function. However, if urine or water has streamed into the pump, discontinue
use and contact Customer Support at 1-888-201-1586.
Although the canister can hold up to 2000cc (mL), the urine should be emptied regularly
from the collection canister before volume reaches 1800cc (mL). Failure to empty
canister before urine overflow may cause damage to the PureWick™ Urine Collection
System and is not covered under the warranty.
It is important that the port connections be connected correctly for proper operation of
the PureWick™ Urine Collection System.
Use of this equipment next to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
Portable radio frequency (RF) communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 12 inches
(30cm) to any part of the PureWick™ Urine Collection System, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
Use only PureWick™ Urine Collection System accessories with this device. Incompatible
parts or accessories can result in degraded performance and will void the warranty.
Do not use accessories past expiration date indicated on package labeling.
Use only the PureWick™ Urine Collection System A/C power cord with the device. Use
of an alternate consumer style A/C power adapter or an extension cord may cause
damage to device and electrical shock or injury and will void the warranty.
Do not place PureWick™ Urine Collection System or its cord across walkways creating a
tripping hazard.
For PW200: The battery cells used in this device may present a fire or chemical hazard.
To minimize the risk of damaging the battery inside the PureWick™ Urine Collection
System, do not use the device outside the indicated operating temperature range of
41°F - 104°F (5°C - 40°C). The battery is not intended to be replaced; replacement could
result in a hazard.
TO REDUCE THE RISK OF ELECTRICAL SHOCK OR INJURY, DO NOT DISASSEMBLE
THIS UNIT. NO MODIFICATION OF THIS EQUIPMENT IS ALLOWED.
KIT CONTENTS
Reusable Items:
• PureWick
Urine Collection System,
TM
with or without battery (1)
• A/C Power Cord (1)
• User Guide (1)
Single-Patient Use Only Items (Accessories) :
• 2000cc (mL) Collection Canister with Lid (1)
• Collector Tubing with Elbow Connector (1)
• Pump Tubing (1)
• Privacy Covers (2)
2