• Cleaning: Prepare a soapy solution by mixing 1 teaspoon (approximately 5 mL)
of dish soap with 1 gallon (approximately 4 L) of cool tap water. Fully submerge
canister and lid in the solution. Allow it to sit for a minimum of ten (10) minutes.
While submerged, use a soft brush (e.g. toothbrush) to brush all accessible areas of
the canister and lid for a minimum of one (1) minute to remove any visible signs of
debris or dirt.
• Rinse: Rinse the canister and the lid thoroughly with cool tap water until there is no
visible sign of the cleaning solution.
• Visual inspection: Inspect the canister and the lid to ensure that all debris and dirt
have been removed. If there is any sign of debris or dirt, repeat steps above until
there is no sign of debris or dirt remaining on the canister and lid.
• Disinfection: Fully submerge the canister and lid in 70% isopropyl alcohol (IPA).
Allow it to sit for a minimum of ten (10) minutes.
• Rinse: Rinse the canister and the lid thoroughly with cool tap water.
• Drying: Dry the canister and lid with a clean low lint towel or cloth.
Collector Tubing (with elbow connector) and Pump Tubing
• Initial Rinse: Disconnect the collector tubing from the elbow connector and
disconnect the elbow connector from the lid. Rinse the disconnected tubing
thoroughly by running cool tap water through the inside of the tubing. While
rinsing, remove any excess dirt by wiping the outside of the tubing and elbow
connector with low lint disposable wipes. While rinsing, flush the tubing with cool
tap water to remove any excess dirt.
• Cleaning: Prepare a soapy solution by mixing 1 teaspoon (approximately 5 mL) of
dish soap with 1 gallon (approximately 4 L) of cool tap water. Completely submerge
the tubing and elbow connector in the solution and allow it to soak for a minimum of
ten (10) minutes. At the beginning of the soak time, flush the inside of the tubing with
the prepared solution. While submerged, use a soft brush (e.g. toothbrush) to brush all
accessible areas of the tubing and elbow connector for a minimum of one (1) minute
to remove any visible signs of debris or dirt.
• Rinse: Rinse the tubing and elbow connector thoroughly with cool tap water until
there is no visible sign of the soap cleaning solution. While rinsing, flush the inside
of the tubing with the tap water.
• Visual inspection: Inspect the tubing to ensure that all debris and dirt have been
removed. If there is any sign of debris or dirt, repeat the steps above until there is
no sign of debris or dirt remaining on or inside the tubing.
• Disinfection: Fully submerge the tubing and elbow connector in 70% isopropyl
alcohol (IPA). Allow it to soak for a minimum of ten (10) minutes. Flush the IPA
through the tubing, prior to starting the ten (10) minute time.
• Rinse: Rinse the tubing and elbow connector thoroughly with cool tap water,
ensuring that the inside of the tubing is flushed with water.
• Drying: Dry the tubing and elbow connector outer surface with a clean low lint
towel or cloth and allow the inside of the tubing to air dry for a minimum of thirty
(30) minutes at room temperature.