Reflex Functions
Signal sources:
Tone - Contra, Reflex:
Tone - Ipsi, Reflex:
NB noise - Contra, Reflex:
NB noise - Ipsi, Reflex:
Noise - Contra, Reflex:
Noise - Ipsi, Reflex:
Outputs:
Contra Earphone:
Ipsi Earphone:
Air:
Test types:
Manual Reflex
Automated Reflex
Reflex Decay
Reflex latency
ABR Infant Screening
Preamplifier
One Channel
Gain:
Frequency response:
Noise:
CMR Ratio:
Max input offset voltage:
Input impedance:
Power from main unit:
Electrical Impedance
Measurement frequency:
measurement
Waveform:
Measurement current:
Range:
Stimulus
Stimuli
Stimulus rate:
Transducers:
Channels:
Level:
Bandwidth:
Recording
Analysis time:
A/D resolution:
Artifact reject system:
Display
Algorithmic Sensitivity
CE-Chirp
Algorithmic Specificity
CE-Chirp
Titan Instruction for Use - English
Date: 2011-06-06
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz.
500, 1000, 2000, 3000, 4000 Hz.
250, 500, 1000, 2000, 3000, 4000, 6000, 8000 Hz.
1000, 2000, 3000, 4000 Hz.
Wide Band, High Pass, Low Pass.
Wide Band, High Pass, Low Pass.
TDH39 earphone, DD45 earphone, CIR55 insert and/or
EARtone 3A insert for Reflex measurements.
Probe earphone incorporated in the probe system for Reflex
measurements.
Connection of the air system to the probe.
Manual control of all functions.
Automatic reflexes:
Automatic, 10 dB above threshold and manually controlled with
stimulus durations of 10 to 30 s.
Automated, first 300 ms from stimulus start.
3 electrodes. 50 cm
Switchable: Software will automatically switch mastoid and
ground if mastoid montage is used. So the user does not need to
replace electrode during testing.
64 dB
0,5 - 5000 Hz
<25 nV/√Hz
>90 dB.
2.5 V
10 MΩ/ 170 pF
Isolated power supply
33 Hz
Rectangular
11.25 μA
0.5 kΩ – 25 kΩ ± 10 %
Click range (177-11313 Hz)
CE-Chirp® range (177 – 11313 Hz)
HiLo CE-Chirp® range
90 Hz
EarTone 3A insert phone auto calibrated & detection TDH 39
head phone auto calibrated & detection
IOW Probe auto calibrated & detection
2
30 dB nHL, 35 dB nHL, 40 dB nHL
22.05 kHz
1-10 min or Residual noise 5-80 nV
24 bit
Rejection level (Peak, Min RMS, Max RMS) Clipping
(Saturation)
Stimulus level and type, Bar and Graph view
Basic and advanced view
®
99.9%
®
> 96%
Page 15/21
-
Single intensities
-
Reflex growth
(Lo 177- 1414 Hz) & (Hi 1414 – 11313 Hz)

Publicidad

Tabla de contenido
loading

Tabla de contenido