Interacoustics Impedance Audiometer Titan Manual De Instrucciones página 25

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 101
Coupler Types used by Calibration
IMP:
TDH39 is calibrated using a 6cc acoustic coupler made in accordance to IEC 60318-3, Ipsilateral
earphone and probe tone are calibrated using a 2cc acoustic coupler made in accordance to IEC
60318-5
DPOAE:
Probe stimuli L1 and L2 are calibrated in SPL values using an ear simulator coupler made in
accordance to IEC 60318-4.
ABRIS:
Probe and insert stimuli are calibrated in SPL values using an ear simulator coupler made in
accordance to IEC 60318-4. Headphones (TDH39 and DD45) stimuli are calibrated in SPL values
using an artificial ear coupler according to IEC
General about Specifications
Interacoustics continuously strives to improve its products and their performance. Therefore the
specifications can be subject to change without notice.
The performance and specifications of the instrument can only be guaranteed if it is subject to a
technical maintenance at least once a year. This should be carried out by a workshop authorized
by Interacoustics.
Interacoustics puts diagrams and service manuals at the disposal of authorized service companies.
Enquiries about representatives and products may be sent to:
Interacoustics A/S
Drejervaenget 8
DK 5610 Assens
Denmark
Titan Instruction for Use - English
Date: 2011-06-06
60318-1.
Phone:
+45 63713555
Fax:
+45 63713522
e-mail:
info@interacoustics.com
http:
www.interacoustics.com
Page 21/21