C3
C1
Connector
adjustment ring
PEP
Diam. Hole
Val. Flow
Ø 2
7,5 l/min
Ø 3
6,4 l/min
Ø 4
3,9 l/min
Ø 5
3,3 l/min
Ø 6
3,2 l/min
C2
C3
A2
C1
mbar
+
I
O
timer
Flow regulator
Adjust the ring nut C1 of the connector C3 and the flow regu-
lator A3 of the device for the desired respiratory physiothera-
- +
py according to the values reported in the following table.
A3
Flow regulator on MAX
Val. Press
7 mbar
6 mbar
3 mbar
2 mbar
1 mbar
Digital pressure gauge
- During the patient's exhalation phase, the pressure gauge
(A2) displays the positive pressure generated according to the
hole selected on the ring nut (C1) of the PEP connector (C3).
- +
- The pressure gauge is only designed to indicate the correct
patient exhalation. It is not an indicator of the precise value
of exhalation flow.
Use the "+ " and "- " keys to set the therapy duration. The thera-
py may last 5, 10, 15 or 20 minutes.
At the end of the set therapy, the 4 LEDs flash and an acoustic
warning beeps 5 times.
Flow regulator set on MIN
Val. Flow
Val. Press
1,2 l/min
2 mbar
1,1 l/min
1 mbar
1,0 l/min
0 mbar
0,8 l/min
0 mbar
0,6 l/min
0 mbar