The power cable (A12) has not been in-
serted properly into the device power
socket (A11) or into the mains socket.
When I inhale, the
The switch (A1) is set to "OFF".
compressor does not
The pressure switch is faulty.
activate and no pulsed
air comes out of the
The air tubes (A9) (A10) are not proper-
mouthpiece.
ly connected to the unit (A) and/or the
PEP connector (C3)
The flow regulator filter (A3) is dirty.
The device (A) is very
The pressure regulator (A3) is not cor-
noisy.
rectly inserted in its housing
If, after checking the aforementioned conditions, the device still does not work properly, we recommend that you
contact your local dealer or an authorised FLAEM service centre.
Insert the power cable correctly into the socket
(A11) and into the relative mains socket.
Set switch (A1) to "ON" or contact a FLAEM ser-
vice centre to replace the pressure switch.
Check that the tubes are correctly connected
to the relative sockets of the device (A). See the
connection diagram on the cover.
Replace the filter (A3).
Properly insert the pressure regulator (A3) in its
housing.